DOTX-45 Dissolved Oxygen System Part 7 –Calibration
DOTX-45 50 -
7.2 Temperature Calibration
The temperature calibration sequence is essentially a 1-point offset calibration
that allows adjustments of approximately ±5 °C.
The sensor temperature may be calibrated on line, or the sensor can be removed
from the process and placed into a known solution temperature reference. In any
case, it is critical that the sensor be allowed to reach temperature equilibrium with
the solution in order to provide the highest accuracy. When moving the sensor
between widely different temperature conditions, it may be necessary to allow the
sensor to stabilize as much as one hour before the calibration sequence is
initiated. If the sensor is on-line, the user may want to set the output HOLD
feature prior to calibration to lock out any output fluctuations.
1. Scroll to the CAL menu section using the MENU key and press ENTER or the
UP arrow key.
2. Press the UP arrow key until Cal Temp is displayed.
3. Press the ENTER key. The message Place sensor in solution then press
ENTER will be displayed. Move the sensor into the calibration reference (if it
hasnt been moved already) and wait for temperature equilibrium to be
achieved. Press ENTER to begin the calibration sequence.
4. The calibration data gathering process will begin. The message Wait will
flash as data is accumulated and analyzed. The °C or °F symbol may flash
periodically if the reading is too unstable.
5. The message Adjust value - press ENTER will be displayed, and the right-
most digit will begin to flash, indicating that the value can be modified. Using
the UP and LEFT arrow keys, modify the value to the known ref solution
temperature. Adjustments up to ± 5 °C from the factory calibrated
temperature are allowed. Press ENTER.
Once completed, the display will indicate PASS or FAIL. If the unit fails, the
temperature adjustment may be out of range, the sensor may not have achieved
complete temperature equilibrium, or there may be a problem with the
temperature element. In the event of calibration failure, it is recommended to
attempt the calibration again immediately.