DOTX-45 Dissolved Oxygen System Part 6 Configuration
DOTX-45 39 -
and therefore, it is best in most cases to leave it at the
default value. Increasing this value will make the controller
more responsive.
AC POWERED UNITS ONLY
*Set 4 mA #2
*Set 20 mA #2
[temp/D.O.] These functions set the second 4 mA and 20 mA current
loop output points for the transmitter. The output may be set
to track temperature (default), PPM oxygen, or % Saturation.
The values stored for the 4 mA point may be higher or lower
than the value stored for the 20 mA point.
The entry value is limited to a value between 0 and 50 °C if it
is set for temperature, within the range specified in Set
Range if the output is set to track oxygen. The 4 mA and
the 20 mA point must be at least 20 units away from each
other. Press ENTER to initiate user entry mode, and the
value will flash. Use arrow keys to modify value. Press
ENTER to store the new value.
NOTE: If the temperature units are changed between °C and
°F (see Temp Units in this section), the default settings for
this output will be stored (present data is not converted.)
ALARM CONFIGURATIONS APPLY TO AC POWERED UNITS ONLY
*A Setpoint This function establishes the oxygen trip point for relay A.
The entry value is limited to a value within the range
specified in Set Range. Use the LEFT arrow key to select
the first digit to be modified. Then use the UP and LEFT
arrow keys to select the desired numerical value. Press
ENTER to store the new value.
*A Hysteresis This function establishes the hysteresis, or deadband, for
Relay A. Hysteresis is most often used to control relay
chattering; however, it may also be used in control schemes
to separate the ON/OFF trip points of the relay. Press
ENTER to initiate user entry mode, and the value will flash.
Use the arrow keys to modify value. Press ENTER to store
the new value.
*A Delay This function places an additional amount of time delay on
the trip point for relay A. This delay is in addition to the main
Vista de pagina 39